Steps
1. Unscrew the cable gland compression nut to relieve the sealing cone and release the cable.
①
②
③
2. Unscrew the probe sleeve from the lower assembly using two 17 mm (0.67") wrenches
across the flat surfaces, as shown below.
①
②
③
– Place two wrenches on the flat surfaces of the probe as shown: hold the probe sleeve
– Remove the lower assembly by turning the threaded electrode end counter-clockwise:
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA
Cable gland compression nut
Probe sleeve
Probe, lower assembly
Flat surface
Flat surface
Lower wrench
still, and turn the lower wrench counter-clockwise to loosen the probe lower assembly.
this exposes the three leads, the tension block, and the steel spacer.
Shortening the cable
F.1 Pointek CLS200 Digital, cable version
139