PROFIBUS PA Parameters
10.1 Measured Values
10.1.2

Changing parameter settings

Note
Initial setup can be carried out prior to mounting into the process, but it is extremely
important to calibrate the unit and adjust the sensitivity on the product itself.
• First launch SIMATIC PDM, connect to Pointek CLS200/300, and upload data from the
device (the status fields change to Loaded).
• Adjust parameter values in the parameter value field.
• After adjusting the value, press Enter (the status fields read Changed).
• When you have completed the adjustments, open the menu Device – Download data to
Device then go to File – Save to save parameter settings offline (the status fields are
cleared).
10.1.3

Filling level status

Adjust 0 % application range setting
The lower limit of the application range (0%) corresponds to the left limit of the bar graph in
the digital display.
For reliable and accurate detection of the process material, select the example in the table
below that most closely resembles your application, and follow the setup procedure.
Application
General applications
Demanding applica-
tions
Interface detection
1. Follow the setup procedure that most closely resembles your application.
2. Open the menu View – Display and select the tab Transducer Block: Discrete Input (Part
1). Make a note of the Sensor_Value (digits)
3. Close Display window, and enter the sensor value (recorded in the previous step) to the
Value field:
the status field reads Changed.
4. If no more settings need adjusting, download the data to the device and save the parameter
settings offline (the status fields go blank). Otherwise, continue to adjust other parameters
as required.
72
Material
dry solids low viscosity liquids
hygroscopic / wet solids high
viscosity and high conductivity
liquids
liquid A / liquid B foam / liquid
> Input > Transducer Block: Discrete Input > Range of Application> 0%. Press Enter:
Setup procedure
Sensor uncovered and a minimum of 100
mm (4") free space all around
Sensor immersed and then uncovered, but
retaining as much build up of material as
possible on the sensor.
Immerse the sensor in the material that
has the lowest relative dielectric constant.
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA