Mode 20: 100% Application Range Setting - Siemens Pointek CLS200 Operating Instructions Manual

Hide thumbs Also See for Pointek CLS200:
Table of Contents

Advertisement

Local operation
7.2 Programming via the Local User Interface (LUI)
7.2.13

Mode 20: 100% Application Range setting

This point corresponds to the left limit of the bar graph in the digital display.
CLS200 100% Application Range setting
To adjust the 100% Application Range setting for CLS200:
1. Install the device in the process.
2. Use the M button to navigate to Mode 20.
3. Set up the process to provide the recommended condition for your process from the table
below:
Application
General
Demanding
Interface detec-
tion
4. Press ⇑ or ⇓. The display switches to the current measured sensor value in counts. Record
the count value.
5. Press ⇑ or ⇓ to enter Edit Mode; the sensor value is frozen and ready for editing.
6. Set the sensor value to approximately 1000 counts below the value obtained in step 4. Press
⇑ to edit the value of each digit, and press ⇓ to move the cursor to the next digit.
7. After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate the
procedure and save the new value. The display now shows the new value for the 100%
Application Range setting.
8. Press M to exit.
48
Material
dry solids
low viscosity liquids
hygroscopic / wet solids
high viscosity and high conductivity liquids
liquid A / liquid B
foam / liquid
Setup conditions
sensor fully covered
sensor fully covered
sensor immersed in the mate-
rial with the highest dielectric
constant
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA