Mounting Restrictions - Siemens Pointek CLS200 Operating Instructions Manual

Hide thumbs Also See for Pointek CLS200:
Table of Contents

Advertisement

Installation
5.1 Pointek CLS200
5.1.3.3

Mounting restrictions

Note
• Keep the sensor at least 50 mm (2") away from any nozzle or tank wall.
• If multiple units are used, allow at least 100 mm (4") between them, to prevent
interference.
In nozzle
18
Close to tank walls
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA

Advertisement

Table of Contents
loading