Adjust 100 % application range setting
Adjusts the upper limit of the application range (100%). This point corresponds to the right
limit of the bar graph in the digital display.
For reliable and accurate detection of the process material, select the example in the table on
page 66 that most closely resembles your application, and follow the setup procedure.
Application
General applications
Demanding applica-
tions
Interface detection
1. Follow the setup procedure that most closely resembles your application.
2. Open the menu View – Display and select the tab Transducer Block: Discrete Input (Part
1). Make a note of the Sensor_Value (digits)
3. Close Display window, and copy the sensor value to the parameter view field: > Input >
Transducer Block: Discrete Input > Range of Application > 100%. Press Enter: the status
field reads Changed.
4. If no more settings need adjusting, download the data to the device and save the parameter
settings offline (the status field goes blank). Otherwise, continue to adjust other parameters
as required.
Inversion
When Inversion Output = On, the level status undergoes a logical inversion. If you want to
invert the logic of the level status:
1. Go to Output > Function Block: Discrete Input > Inversion Output > Off or On
2. Press Enter: the status field reads Changed
Delay
You can adjust the timing behavior of the transmitter using two independent delay times.
• Rise Time (Off to On) determines the delay of the signal flow from the moment the
sensor becomes covered until the Primary Value is set
• Fall Time (On to Off) determines the delay of the signal flow from the moment the
sensor becomes uncovered until the Primary Value is reset.
The delay timers have a range of values from 0.0 to 100.0 seconds.
• Go to Input > Transducer Block: Discrete Input > Delay > Rise Time (Off to On) and set
the value from 0 to 100 seconds.
• Go to Fall Time (On to Off), and set the value from 0 to 100 seconds.
• Press Enter: the status field reads Changed.
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA
Material
dry solids low viscosity liquids
hygroscopic / wet solids high
viscosity and high conductivity
liquids
liquid A / liquid B foam / liquid
PROFIBUS PA Parameters
10.1 Measured Values
Setup procedure
Sensor fully covered
Sensor fully covered
Immerse the sensor in the material that
has the highest relative dielectric con-
stant.
73