5. You can view the reaction in the status of the measured values (input and secondary
variables), and the output, using SIMATIC PDM (open the View Menu Display and select the
appropriate register).
6. To return to normal operation, turn off Simulation Electronic Temperature and press
Transfer.
Drag indicators (Peak indicators)
Two pairs of drag indicators allow you to monitor the negative and positive peak values for
the measured values Sensor Value and Electronics Temperature.
• Open the View Menu Drag Indicators.
• To reset the peaks to the actual measured value, click the appropriate Reset button.
• Press Enter: the status field reads Changed
Operating hours
An operating hours meter for the electronics is activated when the transmitter is first started.
1. Open the View Menu Operating hours.
Switch point adjustment
There are two switch points:
• Switch point 1 (Off to On)
• Switch point 2 (On to Off)
To adjust the switch behavior, set the percentage of the application range
the logic switch will change from OFF to ON, or from ON to OFF.
1. The application range must be defined first (see Adjust 0 % application rangesetting: and
Adjust 100 % application range setting on page 66. Ensure your setup procedure follows the
guidelines for the application which most resembles your operation).
2. Open the menu View–Display, and select the register Transducer Block: Discrete Input, to
view the actual Sensor Value, the Sensor Value Discrete, the actual valid Switch Point and
Hysteresis
3. Go to Input > Transducer Block: Discrete Input > Switch behavior.
4. Edit the value of Switch Point 1 in the "Switch Point" field.
5. Edit the value of Switch Point 2 to adjust the Hysteresis
6. Click on Transmit.
7. Track the effect of the adjustment by looking at the Sensor Value Discrete.
The application range is defined by the 0% and 100% settings (see Adjust 0 % application
1)
range setting).
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA
PROFIBUS PA Parameters
10.1 Measured Values
at which point
1)
77