7. After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate the
procedure and save the new value. The display now shows the new value for the 0%
Application Range setting.
8. Press M to exit.
CLS300 0% Application Range setting
To adjust the 0% Application Range setting for CLS300:
1. Install the device in the process.
2. Use the M button to navigate to Mode 19.
3. Set up the process to provide the recommended condition for your process from the table
below:
Application
General
Demanding
Interface detec-
tion
4. Press ⇑ or ⇓. The display switches to the current measured sensor value in counts. Record
the count value.
5. Press ⇑ or ⇓ to enter Edit Mode; the sensor value is frozen and ready for editing.
6. Set the sensor value to 95% of the value obtained in step 4 (0.95 x counts). Press ⇑ to edit
the value of each digit, and press ⇓ to move the cursor to the next digit.
7. After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate the
procedure and save the new value. The display now shows the new value for the 0%
Application Range setting.
8. Press M to exit.
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA
7.2 Programming via the Local User Interface (LUI)
Material
dry solids
low viscosity liquids
hygroscopic / wet
solids
high viscosity and
high conductivity liq-
uids
liquid A / liquid B
foam / liquid
Setup conditions
Material in tank should be at the lowest point it
could reach during the normal process.
Material in tank should be at the lowest point it
could reach during the normal process. The sensor
should have the minimum possible material build
up.
Material in tank should be at the lowest point it
could reach during the normal process.
Local operation
47