Local operation

7.3 CLS200 digital switching point setup

7.3
CLS200 digital switching point setup
Use the following procedure to set up the switch point for the CLS200 Digital model.
1. Set the 0% Application Range Setting (Mode 19):
– Install the device in the process.
– Use the M button to navigate to Mode 19.
– Set up the process to provide the recommended condition for your process from the
– Press ⇑ or ⇓. The display switches to the current measured sensor value in counts.
– Press ⇑ or ⇓ to enter Edit Mode; the sensor value is frozen and ready for editing.
– Set the sensor value to approximately 300 counts below the value of the current
– After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate
– Press M to exit
2. Set the 100% Application Range setting (Mode 20):
– Install the device in the process.
– Use the M button to navigate to Mode 20.
– Set up the process to provide the recommended condition for your process from the
54
table below:
Application
Material
General
dry solids
low viscosity liquids
Demanding
hygroscopic / wet solids
high viscosity and high conductivity
liquids
Interface detec-
liquid A / liquid B
tion
foam / liquid
reading. Press ⇑ to edit the value of each digit, and press ⇓ to move the cursor to the
next digit.
the procedure and save the new value. The display now shows the new value for the
0% Application Range setting.
table below:
Application
Material
General
dry solids
low viscosity liquids
Demanding
hygroscopic / wet solids
high viscosity and high conductivity
liquids
Interface detec-
liquid A / liquid B
tion
foam / liquid
Setup conditions
sensor uncovered;
min. 100 mm (4") free space
all around
sensor immersed then uncov-
ered; but retaining max. pos-
sible material buildup
sensor immersed in the mate-
rial with the lowest dielectric
constant
Setup conditions
sensor fully covered
sensor fully covered
sensor immersed in the mate-
rial with the highest dielectric
constant
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA