PROFIBUS PA Parameters
10.1 Measured Values
Parameterizable slope
Carry out the following settings:
1. Open the Device Menu Simulation.
2. Select the register Sensor Value.
3. Set Simulation Sensor Value, then set the parameters:
– for simulation mode fixed, enter a Sensor Value
– for simulation mode ramp, adjust the ramp parameters
4. Transfer the settings to the device.
5. You can view the behavior of the Sensor Value Discrete, the Input (Primary) Value, and the
Output, in SIMATIC PDM (open the View Menu Display, and select the register Transducer
Block: Discrete Input Part I).
6. To return to normal operation, turn off Simulation Sensor Value and press Transfer.
Electronics temperature simulation
By simulating the electronics temperature, you can test the effect of excessive temperatures
on the measuring results:
1. Open the Device Menu Simulation
2. Select the register Electronics Temperature.
3. Set Simulation Electronic Temperature, then set the parameters:
– either enter a fixed value for Electronics Temperature fixed
– or adjust the ramp parameters for Electronic Temperature ramp
4. Transfer the settings to the device.
76
Start
End
Value
Number of steps
Time
Step length
Operating Instructions, 05/2021, A5E47904060-AA