4
Place the new o-ring on the Screw Purge Valve.
6
Place the mounting ring on the screw and push down
the o-ring.
1260 SFC Binary Pump
Replace the O-Ring on the Purge Valve
5
Place the screw with o-ring on the Purge Valve Body.
7
Place a new frit into the purge valve with the orientation
of the frit as shown below (slit in frit points to the front).
Reinstall the seal cap including the gold seal.
N O T E
Before reinstallation always check the gold seal in the
seal cap. A deformed seal cap should be exchanged.
8
Maintenance