The firmware of the instrument consists of two independent sections:
• a non-instrument specific section, called resident system
• an instrument specific section, called main system
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232C)
• memory management
• ability to update the firmware of the 'main system'
Main System
Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232C)
• memory management
• ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are
divided into common functions like
• run synchronization through APG remote,
• error handling,
• diagnostic functions,
• or module specific functions like
Firmware Updates
Firmware updates can be done using the following tools (latest version should
be used):
154
internal events such as lamp control, filter movements,
raw data collection and conversion to absorbance.