2 Set the Configuration Switch to DHCP either on the G1369C LAN Interface
Table 18
SW 4
ON
Table 19
SW 6
ON
3 Turn on the module that hosts the LAN interface.
4 Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition,
1260 SFC Binary Pump
Card or the main board of above mentioned modules.
G1369C LAN Interface Card (configuration switch on the card)
SW 5
SW 6
OFF
OFF
LC Modules with 8-bit configuration switch (B-firmware) (configuration switch
at rear of the instrument)
SW 7
SW 8
OFF
OFF
Lab Advisor, Firmware Update Tool) and use MAC address as host name,
e.g. 0030d3177321.
The LC system should become visible in the control software (see Note in
section
"General Information (DHCP)"
Dynamic Host Configuration Protocol (DHCP)
SW 7
SW 8
Initialization Mode
OFF
OFF
DHCP
Initialization Mode
DHCP
on page 179).