Agilent Technologies IDP 7 Instruction Manual

Vpi valve minor replacement kit
Hide thumbs Also See for IDP 7:

Advertisement

Quick Links

IDP 7 / IDP 10
VPI Valve Minor
Replacement
Kit
Manual No. X3807-90015
Rev. A
February 2017
Vacuum Products Division
INSTRUCTION MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies IDP 7

  • Page 1 Vacuum Products Division IDP 7 / IDP 10 INSTRUCTION MANUAL VPI Valve Minor Replacement Manual No. X3807-90015 Rev. A February 2017...
  • Page 2: Documentation Conventions

    IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Preface Documentation Conventions This manual uses the following documentation conventions: Warnings indicate a particular procedure or practice, which WARNING if not followed correctly, could lead to serious injury. Cautions indicate a particular procedure or practice, which if CAUTION not followed, could cause damage to the equipment.
  • Page 3: Installation

    IDP 7 / IDP 10 VPI Valve Minor Replacement Kit IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Part Name Rubber thermal protector Orifice Diaphragm retainer plate Piston O-ring 2-151 O-ring 2-121 O-ring 2-008 Screw Table 1-1 Installation :...
  • Page 4 IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Detach the front cowling from the front of the pump by removing three M8x20mm flange bolts using a 6mm T handled hex key. Bolt Locations Detach the fan cable from the front cowling and remove the cowling from the pump.
  • Page 5 IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Remove the Rear Cowling from the pump assembly. Loosen the three M6x20mm that secure the top part of the base to the Pump Head. Loosen and remove the two M6x16 mm bolts on the bottom of the base to the pump module using a 5mm hex key.
  • Page 6 IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Remove the thermal rubber protector from the inlet assembly and discard. Loosen the four M4x10mm screws holding the inlet flange on to the frame. Remove the inlet flange from the top of the frame assembly. Remove the VPI spring and diaphragm/piston from the frame assembly.
  • Page 7 IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Install a new 10-32 threaded orifice. Apply LVP grease to a new O-ring and install it into the O- ring groove. Install a new O-ring from the kit into the diaphragm/piston assembly. Install the new diaphragm/piston assembly from the kit into the frame.
  • Page 8 IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Install a new Thermal Rubber Protector onto the inlet assembly. Reseat the base assembly. Verify that all connections are in place and that there are no pinched wires. Tighten the five screws that hold the base assembly in place.
  • Page 9 IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Install the six self-tapping screws and two washers that hold the rear cowling to the chassis and front cowling. Tighten rear cowling Tighten rear cowling screws screws and washers. sLocations Attach the fan cable from the front cowling to the motor.
  • Page 10 IDP 7 / IDP 10 VPI Valve Minor Replacement Kit Power the pump up to insure it is functioning correctly.
  • Page 11 Please follow these instructions whenever one of our products needs to be returned. 1) Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to identify all products that have pumped or been exposed to any toxic or hazardous materials.
  • Page 12: Customer Information

    This page intentionally left blank. 4) HEALTH and SAFETY CERTIFICATION AGILENT TECHNOLOGIES CANNOT ACCEPT ANY PRODUCTS CONTAMINATED WITH BIOLOGICAL OR EXPLOSIVE HAZARDS, RADIOACTIVE MATERIAL, OR MERCURY AT ITS FACILITY. Call Agilent Technologies to discuss alternatives if this requirement presents a problem.
  • Page 13: Section 6) Additional Terms

    Shipping Regulations (IATA, DOT, etc.) and carrier requirements. x Customers receiving an Advance Exchange product agree to return the defective, rebuildable part to Agilent Technologies within 15 business days. Failure to do so, or returning a non-rebuildable part (crashed), will result in an invoice for the non-returned/non-rebuildable part.
  • Page 14: Agilent Technologies

    Agilent Technologies Vacuum Product Division United States & Canada India Southeast Asia Agilent Technologies Agilent Technologies Agilent Technologies Vacuum Product Division Vacuum Products Division Vacuum Products Division 205-A, “A” wing of Galleria, South East Asia (SEA) - Alex Ho 121 Hartwell Avenue...

This manual is also suitable for:

Idp 10

Table of Contents