Site Requirements and Specifications
Site Requirements
Power cords
WA R N I N G
Solvents may damage electrical cables.
Bench Space
The module dimensions and weight (see
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm
(3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench
N O T E
rack. This option helps to save bench space as all modules can be placed into one single
stack. It also allows to easily relocate the instrument to another Lab.
Condensation
Condensation within the module
C AU T I O N
Condensation can damage the system electronics.
26
Prevent electrical cables from getting in contact with solvents.
Exchange electrical cables after contact with solvents.
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
Table 1
on page 27) allow you to place