Pump Head Missing; Index Limit - Agilent Technologies Infinity II SFC Binary 1260 User Manual

Table of Contents

Advertisement

6
Error Information
Module Error Messages

Pump Head Missing

Error ID: 2202, 2212
Pump Head Missing: Pump channel A
B: Pump Head Missing: Pump channel B
The pump-head end stop in the pump was not found.
When the pump restarts, the metering drive moves forward to the mechanical
end stop. Normally, the end stop is reached within 20 s, indicated by an
increase in motor current. If the end point is not found within 20 s, the error
message is generated.
Probable cause
1
2

Index Limit

Error ID: 2203, 2213
Index Limit: Pump channel A
B: Index Limit: Pump channel B
The time required by the piston to reach the encoder index position was too
short (pump).
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is reached too fast, the error
message is generated.
Probable cause
1
2
86
Pump head not installed correctly (screws
not secured, or pump head not seated
correctly).
Broken piston.
Irregular or sticking drive movement.
Defective pump drive assembly.
Suggested actions
Install the pump head correctly. Ensure nothing
(e.g. capillary) is trapped between the pump
head and body.
Exchange the piston.
Suggested actions
Remove the pump head, and examine the
seals, pistons, and internal components for
signs of wear, contamination or damage.
Exchange components as required.
Please contact your Agilent service
representative.
1260 SFC Binary Pump