Using the plastic side of the insert tool, insert new seals
into the pump head.
9
Place the pump housing onto the piston housing.
11
Tighten the lock screw.
1260 SFC Binary Pump
Maintenance of a Pump Head without Seal Wash
8
Place the support rings on the piston housing. Note the
correct position of the pins on the support ring.
10
Insert the pistons and carefully press them into the
seals.
12
Install the pump head, see .