Open Cover - Agilent Technologies Infinity II SFC Binary 1260 User Manual

Table of Contents

Advertisement

6
Error Information
General Error Messages

Open Cover

Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.
Probable cause
1
2
3
4
76
The top foam was removed during
operation.
Foam not activating the sensor.
Defective sensor or main board.
Rear of the module is exposed to strong
direct sunlight.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Ensure that the rear of module is not directly
exposed to strong sunlight.
1260 SFC Binary Pump