3
Remove the seal cap from the purge valve. Using a pair
of tweezers or a toothpick, remove the frit.
Valve body
PTFE frit with
groove
Seal cap
5
Insert the purge valve into the purge valve holder.
1260 SFC Binary Pump
Exchange the Purge Valve Frit or the Purge Valve
4
Place a new frit into the purge valve with the orientation
of the frit as shown below (slit in frit points to the front).
Reinstall the seal cap including the gold seal.
6
Tighten the purge valve and reconnect outlet capillary
and waste tubing.