Agilent Technologies Infinity II SFC Binary 1260 User Manual page 178

Table of Contents

Advertisement

12
LAN Configuration
Initialization Mode Selection
Using Default
When Using Default is selected, the factory default parameters are taken
instead. These parameters enable a TCP/IP connection to the LAN interface
without further configuration, see
Figure 21
Using the default address in your local area network may result in network problems. Take
N O T E
care and change it to a valid address immediately.
Table 17
IP address:
Subnet Mask:
Default Gateway
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the module must reside in the same
subnet.
The user may open a Telnet session using the default IP address and change
the parameters stored in the non-volatile memory of the module. He may then
close the session, select the initialization mode Using Stored, power-on again
and establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep
the default parameters to establish the TCP/IP connection.
In the Using Default mode, the parameters stored in the memory of the module are not
N O T E
cleared automatically. If not changed by the user, they are still available, when switching
back to the mode Using Stored.
178
Using Default (Principle)
Using Default Parameters
192.168.254.11
255.255.255.0
not specified
Table 17
on page 178.
1260 SFC Binary Pump

Advertisement

Table of Contents
loading

This manual is also suitable for:

G4782a

Table of Contents