Parameter Indexing
Parameters are indexed when they apply to more than one input or output. The index
value defines the input/output for that parameter. Indexed parameters contain a value for
each index, even if that index is not used.
MultiRanger Display
The index number and the index values are displayed above the parameter indicator on
the LCD.
Notes
•
Transducers are always indexed when the dual point option is enabled.
•
An indexed transducer is commonly referred to as a Point (short for
'Measurement Point'). Point Number refers to indexed transducers.
•
To set all indexed values for a parameter to the same value, use index 0.
•
MR 200 only: Transducer parameters are indexed only if Operation (P001) is set
to Difference (value=4) or Average (value=5) on a single point MultiRanger.
Accessing a Parameter Index
1.
Press DISPLAY
2.
Enter the new parameter number.
3.
Press DISPLAY
4.
Press the number of the required index. Or press ARROW keys
through the available values.
: For optimum performance, set values accurately for indexed parameters.
Note
Ensure that the correct index value is being changed for each parameter value.
7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
Index Number
1
2
3
once to clear current parameter field.
twice.
Index Value
4
5
6
to scroll
Page 35