•
from Transducer One
Send the integer value 01008 to register 46,921.
2.
Read the value from the appropriate parameter register (the example uses 44,921).
Types of values are:
Numeric Values
•
Bit Values
•
Split Values
•
Text Messages
•
Relay Function Codes (P111 Only)
•
A value of 22,222 indicates that an error occurred. Specify a different format type
and try again.
Writing Parameters
The method of writing parameters is similar to the method of reading them. Become
Reading Parameters
familiar with
To write parameter values to the MultiRanger, follow these steps:
Global Index Method (P782 = 0)
1.
Write the primary index value into R43,999.
2.
Write the secondary index value into R43,998.
3.
Write the desired format value into the appropriate format register.
4.
Write the value to the appropriate parameter register.
Parameter Specific Index Method (P782 = 1)
1.
Write the primary index, secondary index, and data format values into the
appropriate format register.
2.
Write the value to the appropriate parameter register.
Page 104
MultiRanger 100/200 – INSTRUCTION MANUAL
on page 107
on page 107
on page 108
on page 109
, page 103, before attempting to write any parameters.
on page 110
7ML19981FB05