mA Output (P200 to P219)

P200 mA Output Range
Determines the mA output range.
Primary Index
Values
Related
If either 1 or 2 is selected, the mA output is directly proportional to the mA Function. If
either 3 or 4 is selected, then the output is inversely proportional.
P201 mA Output Function
Alters the mA output/measurement relationship.
Please note that parameter P201 values vary between the MultiRanger 100 and the
MultiRanger 200.
MultiRanger 100
Primary Index
Values
Related
Altered By
Page 146
MultiRanger 100/200 – INSTRUCTION MANUAL
mA output
0
off
1
0 to 20 mA
2
*
4 to 20 mA
3
20 to 0 mA
4
20 to 4 mA
P911 mA Output Value
mA output
value
mA function
0
OFF
1
level
2
space
3
distance
P202 mA Output Allocation
P911 mA Output Value
P001 Operation
Operation (P001)
level
space
distance
7ML19981FB05