P664 Temperature
View the transducer temperature in °C.
Primary Index
Values
Altered By
Related
Value is displayed when
page 22).
If Temp Source (P660) is set to any value other than Fixed Temp, the value displayed is the
temperature measured. If Temp Source is set to Fixed Temp, the P661 value is displayed.
Rate (P700 to P708)
These parameters determine how material level changes are reported.
P700 Max Fill Rate
Adjusts the MultiRanger response to increases in the actual material level (or advance to
a higher Failsafe Material Level, P071).
Primary Index
Values
Altered by
Related
Enter a value slightly greater than the maximum vessel filling rate. This value, in Units
(P005) or % of Span (P007) per minute, is automatically altered when Maximum Process
Speed (P003) is altered.
P003 Value
1
2
3
7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
Transducer
Range: -50 to 150 (view only)
•
P660 Temp Source
•
P651 Sound Velocity Calibration
•
P653 Velocity
•
P654 Velocity at 20°C
•
P661 Temp Fixed
is pressed in RUN mode (see
Single Point Model
Global
Range: 0.000 to 9999
•
P003 Maximum Process Speed
•
P005 Units
•
P007 Span
•
P071 Failsafe Material Level
0.1
1
10
Readings in Run Mode
Dual Point Model
Transducer
Meters/Minute
on
Page 183