Open Channel Monitoring (OCM)
[MR 200]
An OCM installation is defined one of three ways, based on the Primary Measuring
Device (PMD):
1.
Dimensional (P600 = 2,3,6,7)
For some common weir and flume types. PMD dimensions (P602) are entered directly.
BS-3680 / ISO 1438/1 Thin plate V notch weir on page 74
BS-3680 / ISO 4359 Rectangular Flume on page 75
Palmer Bowlus Flume on page 76
H Flume on page 77
2.
Exponential (P600 = 1)
For most other weir and flume types. PMD exponents provided by the manufacturer are
entered. Flow is calculated using the exponent (P601) and the maximum values (P603 and
P604).
Standard Weirs on page 78
Parshall Flume on page 79
Leoplod Lagco on page 80
Cut Throat Flume on page 81
3.
Universal (P600 = 4,5)
For all other PMDs, the head-to-flow curve can be plotted based on known breakpoints,
usually supplied by the PMD manufacturer.
Typical Flow Characterization on page 82
Example Flumes on page 83
Example Weirs on page 83

Common Parameters

These Quick Start parameters are required for all installations.
Parameter
P001
P002
P003
P004
P005
P006
P007
P801
Page 72
MultiRanger 100/200 – INSTRUCTION MANUAL
Index
Value Description
G
6
Operation
G
1
Material
G
2
Max. Process Speed
G
102
Transducer
G
1
Units
G
1.8
Empty
G
1.0
Span
G
0.8
Range Extension to avoid LOE
= OCM
= liquid
= medium
= XPS-10
= meters
= 1.8m
= 1.4m
7ML19981FB05