Ma Output Limits (P212 And P213) - Siemens MAG 1100 Operating Instructions Manual

Electromagnetic flowmeters
Table of Contents

Advertisement

P210 0/4 mA Output Setpoint
Sets the process level corresponding to the 0 or 4mA value.
Primary Index
Values
Related
Enter the value (in applicable units or %) to correspond to 0 or 4 mA.
P211 20 mA Output Setpoint
Sets the process level that corresponds to the 20 mA value.
Primary Index
Values
Related
Enter the value (in applicable units or %) to correspond to 20 mA.

mA Output Limits (P212 and P213)

Use these features to adjust the minimum and/or maximum mA output values, which
should suit the input limit requirements of the external device.
P212 mA Output Min Limit
Sets the minimum mA output value (in mA) to be produced.
Primary Index
Values
Related
Preset is determined by mA Function (P200). If P200 = 1 or 3, then the preset is 0.0, or if
P200 = 2 or 4, then the preset is 3.8. For P200=1 or 3 (preset is 0.0), this parameter has no
effect because the minimum limit cannot be negative, and the minimum current is always
0.0 mA.
7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
mA output
Range: -999 to 9999
P211 20 mA Output Setpoint
mA output
Range: -999 to 9999
P210 0/4 mA Output Setpoint
mA output
Range: 0.000 to 22.00
Preset: 0.0 or 3.8
P200 mA Output Range
P213 mA Output Max Limit
Page 149