Hardware Information
Setting the 8-bit Configuration Switch (without On-board) LAN

Special Settings

The special settings are required for specific actions (normally in a service
case).
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident mode. It is not operable as a
module. It only uses basic functions of the operating system for example, for
communication. In this mode the main firmware can be loaded (using update
utilities).
Table 18
Boot Resident Settings (without on-board LAN)
Mode Select
TEST/BOOT
188
SW1
SW2
SW3
1
1
SW4
SW5
0
0
1
SW6
SW7
0
0
1260 Infinity QuatPump VL User Manual