Table 11 Equipment Required To Verify Trigger Sensitivity - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

2
Testing Performance
48
Test limits for:
• Internal trigger on all models:
< 10 mV/div: greater of 1 div or 5 mV
>=10 mV/div: 0.6 div
• External trigger on all 2- channel models
(MSO/DSO7xx2A/B):
Trigger range: 1.0V
DC to 100 MHz: < 100 mV
>100 MHz to max bandwidth: < 200 mV
Trigger range: 8.0V
DC to 100 MHz: < 250 mV
>100 MHz to max bandwidth: < 500 mV
• External trigger on 4- channel models:
(MSO/DSO7xx4A/B)
DC to 500 MHz: < 500 mV
Equipment Required to Verify Trigger Sensitivity
Table 11
Equipment
Critical Specifications
Signal
25 MHz, 100 MHz, 350 MHz,
Generator
500 MHz, and 1 GHz sine
waves
Power splitter Outputs differ < 0.15 dB
Cable
BNC, Qty 3
Adapter
N (m) to BNC (f), Qty 3
BNC (f) to BNC (m)
Feedthrough
50
pp
pp
pp
pp
Recommended Model/Part
Agilent N5181A, E4400B, or
8648A
Agilent 11667A
Agilent 10503A
Agilent 1250-0780
Agilent 0960-0301
7000 Series Oscilloscopes Service Guide