Before Testing Performance
Let the oscilloscope warm up before testing.
N O T E
The oscilloscope under test must be warmed up (with the oscilloscope application running)
for at least 30 minutes prior to the start of any performance test.
Equipment Required
Table 15
Description
Digital Multimeter
Cable Assembly
Cable Assembly
Adapter
Adapter
Adapter
Adapter
Calibration
1 Push the Utility button on the front panel.
2 Select Self-Cal menu item in the Utility menu.
3 Follow the on- screen instructions.
Agilent 3000 Series Oscilloscopes User's Guide
Equipment Required for Performance Verification Testing
Critical Specifications
DC voltage measurement accuracy better
than ±0.1% of reading
50Ω characteristic impedance
RS-232 (f)(f)
BNC Barrel (f)(f)
BNC shorting cap
Precision BNC (2)
BNC (f) to dual banana
Service
Recommended
Model/Part Numbers
Agilent 34401A
Agilent 54855-61620
Agilent 34398A
Agilent 1250-0080
Agilent 1250-0929
Agilent 54855-67604
Agilent 1251-2277
129