Equipment Required - Agilent Technologies U3042AE08 User Manual

Table of Contents

Advertisement

U3042AE08
Table 15 N5242A Operation Verification Limit for Reflection Tracking
Frequency
10 MHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 10 GHz
10 GHz to 15 GHz
15 GHz to 20 GHz
20 GHz to 26.5 GHz
1.
Reflection Tracking takes into account Source Loss, Receiver
Loss, Margin, and analyzer's Mixer Cal.
2. A standard unit is one without Options 001 or 002.
NOTE

Equipment Required

The Agilent U3042AE08 requires that the user be familiar with the equipment and
components listed in
This section provides an equipment list and setup of the analyzer and test set.
Table 16
Description
N4691A 3.5 mm ECal Module 10 MHz - 26.5 GHz (Option 00F or M0F) or
N4691B 3.5 mm ECal Module 300 kHz - 26.5 GHz (Option 00F or M0F) or
Mechanical cal kit 85052B or 85052D
N5230C 4-Port Network Analyzer (Option 245 and 551) or
N5242A Option 400 and 551
Set of interconnect cables (analyzer and Test Set), see
Interface Cable Connections" on page
46
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html
2
Standard
Port 1-12
−10.0 dB
−13.0 dB
−16.0 dB
−17.0 dB
−25.0 dB
−30.0 dB
If you suspect that your 12-Port configuration is not operating properly,
ensure that all front RF jumper interconnect cables are correctly attached.
Table
16.
Equipment List
15.
Options
001 or 002
Port 1-12
0 dB
−2.5 dB
−5.0 dB
−7.5 dB
−15.0 dB
−40 dB
"N5230C RF
User's Guide U3042-90002
Operational Check
1
Qty
1
1
1

Advertisement

Table of Contents
loading

Table of Contents