Calibrate Screen
Screen 3 3 3
Calibrate
Calibrate
Screen
Calibrate Screen 3 initiates a calibration of an
accessory solvent meter. During the test, the Volume
Verification screen will appear.
The meter and lines must be primed with
solvent before doing the calibration. See
Solvent Meter Calibration, page 97
instructions.
To initiate the calibration, press the Volume Check
button.
The screen displays the volume dispensed. Enter
the amount of solvent dispensed in the Measured
Volume field, or press
After the Measured Volume is entered, the Accept
Calibration window will appear. Press
the calibration. Press
and retain the previous K-factor.
Press and hold the Reset button
seconds to reset the volume counter.
Figure 71 Calibrate Screen 3
3A4486D
for complete
to end the test.
to accept
to cancel the calibration
for 1-2
Figure 72 Enter Measured Volume of Solvent
Figure 73 Accept Calibration
Setup Mode Screens
89