Frequency Sweep
sweeps using other trigger modes is 250,000 s, and the maximum total sweep time for logarithmic sweeps is 500
s.
Front-Panel:
l
SCPI:
[SOURce[1|2]:]SWEep:TIME {<seconds>|MIN|MAX}
l
Hold Time
Hold time specifies time (in seconds) to remain at the stop frequency, and return time specifies the number of seconds
to return from the stop frequency to the start frequency.
Hold time and return time: 0 to 3600 seconds (default 0).
l
Front-Panel:
l
SCPI:
l
SWEep:HTIMe {<seconds>|MINimum|MAXimum}
SWEep:RTIMe {<seconds>|MINimum|MAXimum}
Marker Frequency
If desired, you can set the frequency at which the signal on the front-panel Sync connector goes to a logic low during
the sweep. The Sync signal always goes from low to high at the beginning of the sweep.
Marker frequency: 1 µHz to 30 MHz (limited to 200 kHz for ramp). The default is 500 Hz.
l
When the sweep mode is enabled, the marker frequency must be between the specified start frequency and stop
l
frequency. If you attempt to set the marker frequency to a frequency not in this range, the instrument will set the
marker frequency equal to the start frequency or stop frequency (whichever is closer).
The
Sync enable setting
l
marker is enabled (with sweep mode also enabled), the Sync setting is ignored.
94
is overridden by enabling the marker used with the sweep mode. Therefore, when the
Agilent 33500 Series Operating and Service Guide