AC Amplitude (high-impedance) Adjustment
Setup
Nominal Signal
DC Level
22
+0.68 V
23*
-0.68 V
24
+1.7 V
25*
-1.7 V
26
+4.3 V
27*
-4.3 V
28
+10.8 V
29*
-10.8 V
30
+0.044V
31*
-0.044V
32
+0.11 V
33*
-0.11 V
34
+0.28 V
35*
-0.28 V
36
+0.68 V
37*
-0.68 V
* Constants are stored after completing this setup.
3. Using the numeric keypad or knob, adjust the displayed voltage at each setup to match the measured voltage.
Select ENTER VALUE.
4. After performing setup 37:
a. If your calibration procedures require you to verify this adjustment, exit the calibration menu and perform
Amplitude (high-impedance) Verification.
b. If you are making all of the adjustments and then verifying the instrument's performance, continue with the
next procedure in this section.