Self-Test Procedures
Error
601
Self-test failed; real time clock settings lost
Real time clock's date-time settings were lost, likely due a disconnected or dis-
charged RTC battery (coin cell found on the front panel board). Error can also
occur if processor board is removed and reinserted into front-panel assembly.
This error condition is captured at power-on, and will be reported by self-test
until the problem is corrected and power is cycled.
602
Self-test failed; main CPU power supply out of range
Processor board detected that one of its supplies was more than 10% out of the
nominal voltage range.
603
Self-test failed; main CPU error accessing boot env
Processor wasn't able to access its boot parameters from flash, possibly due to
out of date firmware, or a problem on the processor board.
604
Self-test failed; front panel processor ping failed
Processor board tried to read front panel revision code and received a 0, possibly
due to un-programmed front panel processor, unseated processor board, or
defective front panel processor.
605
Self-test failed; waveform FPGA not programmed
Processor could not program the waveform FPGA (U1005) at boot-up. The hard-
ware will not work properly.
416
Message and Meaning
Agilent 33500 Series Operating and Service Guide
Probable Cause
RTC Battery
Processor Board
Processor Board
Front Panel Board
Processor Board or Main
Board. See
Trou-
bleshooting
for addi-
tional information on
troubleshooting errors
605-609.