Agilent Technologies Agilent 33500 Series Operating And Service Manual page 182

Waveform generator
Hide thumbs Also See for Agilent 33500 Series:

Advertisement

Parameter
Typical Return
Return the calibration count:
CAL:COUN?
Because the value increments for each calibration point (each CALibration:ALL?), a complete calibration adds
l
many counts.
Can display count regardless of whether instrument is secured.
l
This setting is non-volatile; it will not be changed by power cycling or *RST.
l
CALibration:SECure:CODE < new_code >
Sets the security code to prevent unauthorized calibrations.
Parameter
Unquoted string up to 12 characters
Must start with letter (A-Z)
May contain letters, numbers (0-9) and underscores
Set new security code:
CAL:SEC:CODE MY_CODE_272
When shipped from the factory, the instrument is secured, with the security code set to AT33520A for the
l
33521A and 33522A, or AT33500 for other 33500 Series instruments.
To change code: unsecure calibration memory with old code, then set new code.
l
If you forget the security code, see
l
This setting is non-volatile; it will not be changed by power cycling or *RST.
l
CALibration:SECure:STATe {ON|1|OFF|0} [,< code >]
CALibration:SECure:STATe?
Unsecures or secures the instrument for calibration. To calibrate, you must unsecure the instrument with the code
(CALibration:SECure:CODE).
Parameter
{ON|1|OFF|0}, default ON
<code> is unquoted string up to 12 characters
Unsecure calibration:
CAL:SEC:STAT OFF,MY_CODE_272
Secure calibration:
CAL:SEC:STAT ON
<code> is optional to secure the instrument, but must be correct if provided.
l
Front panel and remote interface calibration use same code. If you secure the instrument from one interface, use
l
Agilent 33500 Series Operating and Service Guide
Typical Return
(none)
Unsecure Instrument Without Security
Typical Return
0 (OFF) or 1 (ON)
CALibration Subsystem
Code.
181