time from the 50% threshold of the rising edge to the 50% threshold of the next falling edge.
4. Set the edge time for both edges.
Press the Edge Time softkey and then set the edge time for both the leading and trailing edges to 50 ns. The edge
time represents the time from the 10% threshold to the 90% threshold of each edge.
Select a Stored Arbitrary Waveform
There are nine built-in arbitrary waveforms stored in non-volatile memory. They are Cardiac, D-Lorentz, Exponential
Fall, Exponential Rise, Gaussian, Haversine, Lorentz, Negative Ramp, and Sinc.
This procedure select the built-in "exponential fall" waveform from the front panel. For information on creating a cus-
tom arbitrary waveform, refer to
1. Select the arbitrary waveform function.
Press the Waveforms button and choose the Arb softkey and then the Arbs softkey.
2. Then choose Select Arb and use the knob to select Exp_Fall. Press Select.
Use Built-in Help System
The built-in help system provides context-sensitive help on any front-panel key or menu softkey. A list of help topics is
also available to assist you with several front-panel operations.
View the help information for a function key
Press and hold down any softkey or button, such as Waveforms . If the message contains more information than will
fit on the display, press the down arrow softkey or use the knob to view the remaining information.