MMEMory Subsystem
Parameter
Both files may be any valid file name
Move the state file MySetup.sta from the currently selected default directory to the folder "Backup" on the internal
flash file system;
MMEM:MOVE "MySetup.sta","INT:\Backup"
Rename the arbitrary waveform arbMonday on the USB drive to the name arbTuesday.
MMEM:MOVE "USB:\arbMonday", "USB:\arbTuesday"
To simply rename a file, specify the same folder for <file1> and <file2>.
l
To move a file to a file of the same name in a different folder, you can specify just the <drive>:<path> for <file2>.
l
MMEMory:STORe:DATA[1|2]  <filename>
Stores the specified arb segment(.arb/.barb) or arb sequence (.seq) data in the channel specified volatile memory
(default, channel 1) in INTERNAL or USB memory.
Parameter
Any valid file name
Store an arbitrary sequence loaded in volatile memory on channel 1 to the internal drive. MMEM:STOR:DATA
"INT:\Sequence1.seq"
When you store an arbitrary waveform segment or sequence (MMEMory:STORe:DATA[1|2]), the instrument's cur-
l
rent settings (voltage values, sample rate, filter type, and so on) are stored in the segment or sequence file. When
you play the file for the first time with FUNCtion:ARBitrary, these settings are loaded and override the instru-
ment's current settings. If you have manually edited a segment or sequence file such that the instrument settings
have been removed, the instrument settings will not be changed when you execute FUNCtion:ARBitrary.
If the destination file exists, it is overwritten, unless marked as hidden or system.
l
If an arbitrary waveform sequence file (.seq) is specified, all the arbitrary waveform segment file names associated
l
with the sequence are stored in the file.
Command will error if the specified arbitrary waveform segment or arbitrary waveform sequence is not found in vol-
l
atile memory.
MMEMory:UPLoad? < filename >
Uploads the contents of a file from the instrument to the host computer.
Parameter
Any valid file name
The following command uploads the contents of the state file "Myfile.sta" in the root directory of the internal flash
file system to the host computer:
MMEM:UPL? "INT:\Myfile.sta"
258
Typical Return
(none)
Typical Return
(none)
Typical Return
IEEE 488.2 definite-length block