Load an arbitrary waveform from the USB drive and set up the balance gain (channel 2 down by 1.02%) and balance
offsets (120 mV for channel 1 and -38 mV for channel 2):
FUNC:ARB "USB:\qam32.barb"
FUNC:ARB:BAL:GAIN 1.02
FUNC:ARB:BAL:OFFS1 0.12
FUNC:ARB:BAL:OFFS2 -0.038
FUNC:ARB:BAL ON
This feature can be used to compensate for minor variations in the load impedances on the two channels, or for
l
minor differences between the two channels of information in the original dual arbitrary waveform file.
When the <state> is on, both channels must stay on the same amplifier range.
l
This setting is stored as part of the instrument state, and it is affected by *RST.
l
FUNCtion:ARBitrary:BALance:GAIN {< percent >|MAXimum|MINimum}
FUNCtion:ARBitrary:BALance:GAIN? {MAXimum|MINimum}
This command requires the IQ Player option.
Sets the gain balance ratio for dual arbitrary waveforms.
-30 to +30
Load an arbitrary waveform from the USB drive and set up the balance gain (channel 2 down by 1.02%) and balance
offsets (120 mV for channel 1 and -38 mV for channel 2):
FUNC:ARB "USB:\qam32.barb"
FUNC:ARB:BAL:GAIN 1.02
FUNC:ARB:BAL:OFFS1 0.12
FUNC:ARB:BAL:OFFS2 -0.038
FUNC:ARB:BAL ON
Both channels must stay on the same amplifier range.
l
A value of 0 means that both channels use their current amplitude. A positive value means that the amplitude of
l
channel 2 is pushed down, while the amplitude of channel 1 remains constant, a negative value pushes channel 1
down while holding channel 2 constant. See the following table for examples.
<percent> Channel 1
Amplitude %
-20
80
-10
90
0
100
10
100
l
Agilent 33500 Series Operating and Service Guide
Parameter
Parameter
Channel 2
Amplitude %
100
100
100
90
Typical Return
Typical Return
+1.02000000E+000
FUNCtion Subsystem
225

Advertisement

loading