STATus:PRESet
Clears
Questionable Data
Parameter
Typical Return
(none)
(none)
Clear enable register bits:
STAT:PRES
STATus:QUEStionable:CONDition?
Queries the condition register for the
Parameter
Typical Return
(none)
+512
Read the condition register (bit 9 is set):
STAT:QUES:COND?
The Questionable Data register group provides information about the instrument's quality or integrity.
l
Any or all conditions can be reported to the Questionable Data summary bit through the enable register.
l
Register is read-only; bits not cleared when read.
l
The condition register bits reflect the current condition. If a condition goes away, the corresponding bit is cleared.
l
*RST
clears the condition register.
l
The query reads the condition register and returns a decimal value equal to the binary-weighted sum of all bits set
l
in the register. For example, if bit 12 (decimal value = 4096) is set, the query returns "+4096".
STATus:QUEStionable:ENABle < enable_value >
STATus:QUEStionable:ENABle?
Enables bits in the
enable register
Status Byte.
Decimal value equal to the sum of the bit decimal values in the register.
Enable bit 9 (value 512) in the enable register:
STAT:QUES:ENAB 512
Use <enable_value> to specify which bits are reported to the Status Byte. The specified value corresponds to the
l
binary-weighted sum of the register bits to enable. For example, to enable bit 5 (value 32) and bit 9 (value 512),
the decimal value would be 544.
Agilent 33500 Series Operating and Service Guide
enable register and
Standard Operation
Questionable Data Register
for the
Questionable Data Register
Parameter
STATus Subsystem Introduction
enable register.
group.
group. The selected bits are then reported to the
Typical Return
+512
289