LAN Configuration
SYSTem:COMMunicate:LAN:GATeway "< address >"
SYSTem:COMMunicate:LAN:GATeway? [{CURRent|STATic}]
Assigns a default gateway for the instrument. The specified IP Address sets the default gateway which allows the
instrument to communicate with systems that are not on the local subnet. Thus, this is the default gateway where
packets are sent which are destined for a device not on the local subnet, as determined by the Subnet Mask setting. If
DHCP is enabled (SYSTem:COMMunicate:LAN:DHCP ), the specified default gateway is not used. However, if the DHCP
server fails to assign a valid IP address, the currently configured default gateway will be used. Contact your LAN admin-
istrator for details.
If you change this setting, you must send
setting.
Parameter
Command: "nnn.nnn.nnn.nnn", default "0.0.0.0"
Query: {CURRent|STATic}, default CURRent
Set default gateway address:
SYST:COMM:LAN:GATEWAY "198.105.232.1"
SYST:COMM:LAN:UPD
CURRent: read address currently being used by the instrument.
l
STATic: read static address from non-volatile memory. This address is used if DHCP is disabled or unavailable.
l
Set to "0.0.0.0" when instrument is shipped from factory or after SYSTem:SECurity:IMMediate.
l
SYSTem:COMMunicate:LAN:HOSTname "< name >"
SYSTem:COMMunicate:LAN:HOSTname? [{CURRent | STATic}]
Assigns a hostname to the instrument. A hostname is the host portion of the domain name, which is translated into
an IP address. If Dynamic Domain Name System (DNS) is available on your network and your instrument uses DHCP,
the hostname is registered with the Dynamic DNS service at power-on. If DHCP is enabled (SYS-
Tem:COMMunicate:LAN:DHCP ), the DHCP server can change the specified hostname.
Parameter
String of up to 15 characters.
Must start with letter (A-Z)
May contain letters, numbers (0-9), or dashes ("-")
Define a hostname:
SYST:COMM:LAN:HOST
Set to "A-33521A-nnnnn" or "A-33522A-nnnnn", where nnnnn is the last five digits of the instrument's serial
l
number, when the instrument is shipped from the factory or after SYSTem:SECurity:IMMediate.
310
If you change this setting, you must send
activate the new setting.
SYSTem:COMMunicate:LAN:UPDate
Typical Return
"198.105.232.1"
SYSTem:COMMunicate:LAN:UPDate
Typical Return
"LAB1-33522A"
Agilent 33500 Series Operating and Service Guide