Disassembly
3. Pull off the instrument bumpers.
4. Loosen the two captive screws (circled in red, below) in the rear bezel and remove the rear bezel.
5. Slide off the instrument cover.
Many of the service procedures can now be performed without further disassembly. Troubleshooting and service pro-
cedures that require power be applied can be performed with the instrument in this state of disassembly.
SHOCK HAZARD. Only service-trained personnel who are aware of the hazards involved should
remove the instrument covers. Dangerous voltages may be encountered with the instrument cov-
ers removed.
422
Agilent 33500 Series Operating and Service Guide