[SOURce[1|2]:]AM:SOURce
[SOURce[1|2]:]AM:SOURce {INTernal|EXTernal|CH1|CH2}
[SOURce[1|2]:]AM:SOURce?
[SOURce[1|2]:]BPSK:SOURce {INTernal|EXTernal}
[SOURce[1|2]:]BPSK:SOURce
[SOURce[1|2]:]FM:SOURce {INTernal|EXTernal|CH1|CH2}
[SOURce[1|2]:]FM:SOURce
[SOURce[1|2]:]FSKey:SOURce {INTernal|EXTernal}
[SOURce[1|2]:]FSKey:SOURce
[SOURce[1|2]:]PM:SOURce {INTernal|EXTernal|CH1|CH2}
[SOURce[1|2]:]PM:SOURce
[SOURce[1|2]:]PWM:SOURce {INTernal|EXTernal|CH1|CH2}
[SOURce[1|2]:]PWM:SOURce?
Select the source of the modulating signal.
{INTernal|EXTernal|CH1|CH2}, default INTernal. BPSK and FSKey can-
not accept CH1 or CH2
Select external modulation source:
AM:SOUR EXT (could also substitute FM, BPSK, FSK, PM, or PWM for AM)
Remarks
If you select EXTernal, the carrier waveform is modulated with an external waveform. Specifically:
l
AM:The modulation depth is controlled by the ±5 V signal level on the rear-panel Modulation In connector.
l
For example, if modulation depth (AM[:DEPTh]) is 100%, then when the modulating signal is at +5 V, the out-
put will be at the maximum amplitude. Similarly, a -5 V modulating signal produces output at minimum ampli-
tude.
FM:If you select the External modulating source, the deviation is controlled by the ±5 V signal level on the
l
rear-panel Modulation In connector.For example, if the frequency deviation is 100 kHz, then a +5 V signal
level corresponds to a 100 kHz increase in frequency.Lower external signal levels produce less deviation and
negative signal levels reduce the frequency below the carrier frequency.
PM:With the External modulating source, deviation is controlled by the ±5 V signal level on the rear-panel
l
Modulation In connector. For example, if you have set the frequency deviation to 180 degrees, then a +5 V
signal level corresponds to a +180 degree phase deviation. Lower external signal levels produce less deviation,
and negative signal levels produce negative deviation.
160
Parameter
Typical Return
INT, EXT, CH1, or CH2
Agilent 33500 Series Operating and Service Guide