STATus Subsystem Introduction
4
Channel 2 Initiated
5
Waiting for Trigger,
Channel 1
6
Waiting for Trigger,
Channel 2
7
(Reserved)
8
Configuration
Changed Event
9
(Reserved)
10
Instrument Locked
11
(Reserved)
12
(Reserved)
13
Global Error
14-
(Reserved)
15
Questionable Data Register Group
The following table describes the Questionable Data Register group.
Bit
Bit Name
0
Channel 1 Volt-
age Overload
1
Channel 2 Volt-
age Overload
2
(Reserved)
3
(Reserved)
4
(Reserved)
5
Loop Unlocked
286
16
32
Instrument is waiting for a trigger. In INIT[1|2]:CONT
OFF, this bit is set after receiving an INIT and while wait-
ing for a trigger. It is cleared after receiving the trigger.
64
This bit is 0 if the channel is in INIT[1|2]:CONT ON
mode.
128
(Reserved for future use)
256
This bit is always 0 in the condition register, as it
reflects an event, not a condition.
512
(Reserved for future use)
1024
If a remote interface (USB or LAN) has a lock (SYS-
Tem:LOCK:REQuest?), this bit will be set. When a
remote interface releases the lock (SYS-
Tem:LOCK:RELease), this bit will be cleared.
2048
(Reserved for future use)
4096
(Reserved for future use)
8192
This is set if any remote interface has an error in its
error queue, and cleared otherwise.
16,384 -
(Reserved for future use)
32,768
Decimal
Value
1
Voltage overload on channel 1 output connector. The out-
put has been disabled.
2
Voltage overload on channel 2 output connector. The out-
put has been disabled.
4
(Reserved for future use)
8
(Reserved for future use)
16
(Reserved for future use)
32
Function generator has lost phase lock. Frequency accuracy
will be affected.
Definition