-360 to +360 degrees, or -2π to +2π radians, as specified by UNIT:ANGLe. Default 0.
Set starting burst phase to 60 degrees:
UNIT:ANGLE DEG
BURS:PHAS 60
Note that BURSt:PHASe is used instead of output phase, and when burst is enabled, the output phase is set to 0.
l
For sine, square, and ramp, 0 degrees is the point at which the waveform crosses 0 V (or DC offset) in a positive-
l
going direction. For arbitrary waveforms, 0 degrees is the first waveform point. Start phase has no effect on noise.
For arbitrary waveforms, BURSt:PHASe is only available if the waveform is 1,000,000 points or less.
l
Start phase also used in gated burst mode. When the gate signal goes false, the current waveform cycle finishes,
l
and output remains at the voltage level of the starting burst phase.
[SOURce[1|2]:]BURSt:STATe {ON|1|OFF|0}
[SOURce[1|2]:]BURSt:STATe?
Enables or disables burst mode.
Parameter
{ON|1|OFF|0}, default OFF
Enable burst mode:
BURS:STAT ON
Output phase is set to 0 when burst is enabled.
l
To avoid multiple waveform changes, enable the burst mode after configuring the other burst parameters.
l
The instrument will not allow the burst mode to be enabled at the same time that sweep or any modulation mode
l
is enabled. When you enable burst, the sweep or modulation mode is turned off.
Agilent 33500 Series Operating and Service Guide
Parameter
Typical Return
0 (OFF) or 1 (ON)
BURSt Subsystem
Typical Return
+6.0000000000000E+01
179