VOLTage Subsystem Introduction
conflict" error.
The high limit sets the highest output voltage allowed to be set, including DC Offset and peak amplitude. It is set in
l
reference to the current
ment's output, then the output limit may not represent the actual voltages at the output connector. For example,
if the output impedance is set to 50 Ω, but the actual load is high impedance, then the actual output peak voltage
may be up to twice the specified limit voltage.
Specifying Voltage Units: You can set the output limit voltage only in volts.
l
When
VOLTage:COUPle[:STATe]
l
affect maximum amplitude and offset voltage settings on both channels. The most restrictive combination of high
and low limits from either channel is used.
[SOURce[1|2]:]VOLTage:LIMit:STATe {ON|1|OFF|0}
[SOURce[1|2]:]VOLTage:LIMit:STATe?
Enables or disables output amplitude voltage limits.
Parameter
{ON|1|OFF|0}, default OFF
Set and enable ±2.5 V output limits on channel 1:
VOLT:LIM:HIGH 2.5
VOLT:LIM:LOW -2.5
VOLT:LIM:STAT ON
When this is turned ON, if the present settings of amplitude and offset exceed the limits, then the limits will be dis-
l
abled. The instrument will generate either a "Settings conflict" error.
When
VOLTage:COUPle[:STATe]
l
affect maximum amplitude and offset voltage settings on both channels. The most restrictive combination of high
and low limits from either channel is used.
Limits are set in reference to the current setting of OUTPut[1|2]:LOAD. If the specified LOAD impedance is not
l
present at the instrument's output, then the output limit may not represent the actual voltages at the output con-
nector. For example, if the output impedance is set to 50 Ω, but the actual load is high impedance, then the actual
output peak voltage may be up to twice the specified limit voltage.
[SOURce[1|2]:]VOLTage:OFFSet {< offset >|MINimum|MAXimum}
[SOURce[1|2]:]VOLTage:OFFSet? [{MINimum|MAXimum}]
Sets DC offset voltage.
Parameter
± 5 VDC into 50 Ω, default 0
Set offset voltage to 100 mV:
VOLT:OFFS 100 mV
The relationship between offset voltage and output amplitude is shown below.
l
324
OUTPUT[1|2]:LOAD
setting. If the specified LOAD impedance is not present at the instru-
is ON, and
VOLTage:LIMit:STATe
Typical Return
0 (OFF) or 1 (ON)
is ON, and
VOLTage:LIMit:STATe
Typical Return
+1.0000000000000E-01
is ON, voltage limit settings on both channels
is ON, voltage limit settings on both channels