AC Amplitude (high-impedance) Adjustment (Channel 2)
The instrument stores a calibration constant for each high-impedance attenuator path. Each path's gain coefficient is
calculated using two measurements: one with the waveform DAC at + output and one with waveform DAC at – out-
put. The setups, therefore, must be performed in pairs.
1. Connect the DMM to the channel output as shown below.
2. Use the DMM to measure the DC voltage at the front-panel connector for each setup in the following table.
Nominal Signal
Setup
DC Level
57
+0.0028 V
58*
- 0.0028 V
59
+0.007 V
60*
-0.007 V
61
+0.017 V
62*
-0.017 V
63
+0.044 V
64*
-0.044 V
65
+0.11 V
66*
-0.11 V
67
+0.28 V
68*
-0.28 V
69
+0.68 V
Agilent 33500 Series Operating and Service Guide
AC Amplitude (high-impedance) Adjustment (Channel 2)
Output of -72 dB range
Output of -72 dB range
Output of -64 dB range
Output of -64 dB range
Output of -56 dB range
Output of -56 dB range
Output of -48 dB range
Output of -48 dB range
Output of -40 dB range
Output of -40 dB range
Output of -32 dB range
Output of -32 dB range
Output of -24 dB range