INITiate Subsystem
Parameter
Change both channels to the wait-for-trigger state:
INIT:IMM:ALL
The trigger system is armed by INITiate[:IMMediate]. Once the trigger count is satisfied, the trigger system
l
returns to idle state and ignores further triggers. The triggered function will be left in whatever state is achieved
with the count of triggers. Rearming the trigger system with another
to apply.
Use
ABORt
to return instrument to idle.
l
If the specified channel has
l
will be generated.
238
Typical Return
INIT:CONT
set ON, these commands have no effect on trigger system and error -213
INITiate[:IMMediate]
Agilent 33500 Series Operating and Service Guide
allows further triggers