SYSTem:COMMunicate:GPIB:ADDRess {< address >}
SYSTem:COMMunicate:GPIB:ADDRess?
Assigns instrument's GPIB (IEEE-488) address, which is displayed at power-on. Each device on the GPIB interface
must have a unique address.
Parameter
Typical Return
0 to 30, default 10
+15
Set GPIB address to 15:
SYST:COMM:GPIB:ADDR 15
Your computer's GPIB interface card has its own address. Avoid using this address for any instrument on the GPIB
l
bus.
This setting is non-volatile; it will not be changed by power cycling or *RST.
l
SYSTem:SECurity:IMMediate
l
Must cycle power for this command to take effect.
l
SYSTem:DATE < yyyy >, < mm >, < dd >
SYSTem:DATE?
Sets system clock date.
Parameter
<yyyy> 2000 to 2100
<mm> 1 to 12
<dd> 1 to 31
Set system date to July 26, 2011:
SYST:DAT 2010, 7,26
SYSTem:ERRor?
Reads and clears one error from error queue.
Parameter
(none)
-113,"Undefined header"
Read and clear first error in error queue:
SYST:ERR?
Up to 20 command syntax or hardware errors can be stored in the error queue.
l
Error retrieval is first-in-first-out (FIFO), and errors are cleared as you read them. The instrument beeps once
l
each time an error is generated (unless disabled by
Agilent 33500 Series Operating and Service Guide
sets GPIB address to 10.
Typical Return
2011,07,26
Typical Return
SYSTem:BEEPer:STATe
SYSTem Subsystem
OFF).