Pressura RPM10 Operation And Service Manual page 97

Room pressure monitor
Hide thumbs Also See for RPM10:
Table of Contents

Advertisement

RPM10 PresSura Monitor
Device
Object Type
Instance
Multi-State
5
Value
Multi-State
6
Value
Multi-State
7
Value
Multi-State
8
Value
Multi-State
9
Value
Multi-State
10
Value
Multi-State
11
Value
Multi-State
12
Value
Multi-State
16
Value
Multi-State
17
Value
Multi-State
18
Value
*The units are based on the value of the Units Value object. When the Units Value is set to 1 or 3, the units are in
English form. When the Units Value is set to 2 or 4, the units are metric. English is the default value.
**The device instance is 010000, summed with the MAC address of the device.
®
BACnet
MS/TP Object Set
*Units
Description
Input 2
Configuration
Input 3
Configuration
Input 4
Configuration
Input 5
Configuration
Input 6
Configuration
Input 7
Configuration
Room 1 Mode
ACH Duct
Status Index
Device Type
Units Value
Writable
Object
Value
4
1
2
3
4
7
1
2
3
1
3
6
2
3
4
5
8
Y
1
2
3
Y
1
2
3
1
2
3
4
5
6
7
20
3
Y
1
2
3
Notes and Range
None
Supply Pressure Flow
Supply Linear Flow
Supply Venturi Flow
Supply Switch
None
Room 1 Door Switch
Room 1 Occupancy
Sensor
None
Room 1 Key Switch
None
None
Exhaust Pressure Flow
Exhaust Linear Flow
Exhaust Venturi Flow
Exhaust Switch
None
Positive
Negative
No Isolation
Supply
Exhaust
Off
Normal
Room 1 Negative Low
Alarm
Room 1 Negative High
Alarm
Room 1 Positive Low
Alarm
Room 1 Positive High
Alarm
Low Exhaust Alarm
Low Supply Alarm
Data Error
RPM10
in H
O, cfm
2
Pa, lps
3
Pa, m
/hr
89