Pressura RPM10 Operation And Service Manual page 46

Room pressure monitor
Hide thumbs Also See for RPM10:
Table of Contents

Advertisement

Input2 Config Menu
TSI Sensor
SOFTWARE
MENU ITEM
NAME
Check Sensor
Check Status
Status
RPM20
ITEM DESCRIPTION
The Check Status item is used to check the
communication status of the sensor. After pressing the
button, the PresSura unit will respond with:
COMM ERROR - DIM cannot communicate with
sensor. Check all wiring and the
pressure sensor address.
SENS ERROR - Physical damage to pressure sensor
circuitry. Unit is not field-repairable.
Send to TSI for repair.
CAL ERROR
- Calibration data lost. Send to TSI for
calibration.
DATA ERROR - Problem with sensor EEPROM, field
calibration or analog output. Check
all data configured and confirm unit
is functioning correctly.
DEFAULT
ITEM RANGE
VALUE
None
N/A