Wiring Information; Back Panel Wiring; Appendix C - Pressura RPM10 Operation And Service Manual

Room pressure monitor
Hide thumbs Also See for RPM10:
Table of Contents

Advertisement

Appendix C

Wiring Information

Back Panel Wiring

Input / Output
PIN #
/ Comm
1, 2
Input
3, 4
Output
5, 6
Input
7, 8
Comm
9, 10
Output
11, 12
Output
13, 14
Input
15, 16
Input
17, 18
Input
19, 20
Input
21, 22
Input
23, 24
Input
25, 26
Output
27, 28
Output
29, 30
Output
31, 32, 33
Comm
34, 35, 36
Comm
Signal
Description
24 VAC/DC
Power in Digital Interface Module (DIM).
24 V
Power for TSI Pressure Sensors 24 VAC
0 to 10 VDC
Input 1
RS-485
Communications between DIM and TSI Pressure
Sensors
Open / Closed
Relay 1 Output (Low Alarm)
Open / Closed
Relay 2 Output (High Alarm or Room Mode)
0 to 10 VDC
Input 2
0 to 10 VDC
Input 3
Open / Closed
Open / Closed
Input 4
0 to 10 VDC
Input 5
Resistance
Resistance
Input 6
Open / Closed
0 to 10 VDC
Input 7
Resistance
0 to 10 VDC
Analog Out 1
0 to 10 VDC
Analog Out 2
4-20 mA
0 to 10 VDC
Analog Out 3
4-20 mA
RS-485
Nurse Station Display
Modbus / Bacnet
BAS Communications
MS/TP / LON
31: B
32: A
33: Ref
34: B
35: A
36: Ref (Modbus / BAcnet MS/TP only)
95